MSI Panel Considerations
Overview
Microsatellite Instability (MSI) is scored from the length distributions of microsatellite (homopolymer) sites. These sites are often in non-coding regions. Thus, a Comprehensive Genomic Profiling (CGP) panel that captures only coding regions of targeted genes will not reliably assess MSI. You must include a dedicated set of microsatellite marker loci in the capture design.
DRAGEN MSI is intended for research use only and is not for use in diagnostic procedures.
Factors Impacting MSI Locus Selection
Microsatellite marker loci are often selected independently of the gene content. This section lists design considerations below citing specifications used by Illumina internal panels. These are designed to provide some initial guidance. DRAGEN is an RUO software. Customers are responsible for the design of their panels and for validating them appropriately.
Design considerations:
Site count — Some Illumina panels assess 130 homopolymer sites with at least 40 qualified sites required for a valid score.
MSI Site Selection — MSI loci should be chosen for signal-to-noise quality, not for proximity to cancer genes. Many are non-coding homopolymer regions. The following characteristics can impact MSI site selection:
Repeat length — MSI sites vary in length. Internal testing suggests using a variety of site lengths can be beneficial.
Germline Variants — In general, sites with prevalent germline variants should be avoided to prevent CNV variants in heterozygous individuals from being interpreted as instability.
Historical Precedence — The most important factor for accurate MSI determination is the inclusion of a sufficient number of MSI sites, not the inclusion of specific sites. Depending on your validation strategy, however, it may be helpful to include the same sites as an orthogonal assay that you will be validating against.
Coverage Targets
MSI marker loci require an intermediate design depth — higher than the HRD scaffold but lower than the gene content.
Gene content (exonic)
≥500×
Variant-calling sensitivity
MSI marker loci
≥60× spanning reads per site.
Reliable homopolymer length distributions.
Note that the number of reads spanning a homopolymer site will be lower than the total depth at that site
HRD scaffold
30–100×
Genome-wide ASCN segmentation
Each microsatellite site must have at least 30 full-spanning reads to be evaluable by default. Sites below the coverage threshold are excluded from analysis. Thus we recommend targeting a higher coverage than this level (60 in the table above).
Recommended Design Procedure
A full list of MSI loci assessed by default in the DRAGEN MSI algorithm is provided on the product files site (see WGS Microsatellite Files). These files have already been filtered for common germline variants. A recommended approach to panel design is:
Intersect the provided MSI sites file with your existing design to see how many MSI sites your panel already covers
Add probes to cover additional sites until you have at least 130 homopolymer sites spanning different genomic regions and length distributions
Create a custom MSI sites list containing the MSI sites targeted in your panel for use in DRAGEN
Verify panel performance (sensitivity and specificity). See Validation Guidance
Command-Line Configuration
DRAGEN MSI requires a microsatellite site file appropriate to the panel and a coverage threshold. Confirm option names and values against the Microsatellite Instability module documentation.
--msi-coverage-threshold
Minimum spanning-read coverage for a microsatellite site. Sites below this are excluded. The default value in tumor-only mode (recommended mode) is 30 reads
--msi-microsatellites-file
Microsatellite site file, intersected with the panel design.
Validation Guidance
It is recommended to confirm:
MSI QC Status
Sensitivity
Specificity
Depending on the intended use case, an accuracy study against an orthogonal method (such as IHC or a ddPCR-based method) may also be warranted.
Confirming MSI QC Status
Below is a typical procedure for verifying expected MSI QC Status.
Process a set of samples (can be microsatellite instability-high (MSI-H) or microsatellite-stable (MSS); can overlap with sample sets used for below tests) that span the expected range of sample quality (e.g., measured by delta Cq, DV200, etc.), coverage, and insert size
Evaluate the range of MSI Sites Assessed across samples to ensure that your expected failure rate is sufficiently small. For some Illumina panels, we consider a sample as an MSI QC failure if fewer than 40 sites are assessed
Note that the following factors can influence MSI Sites Assessed:
Sequencing Coverage
Duplication Rate
Insert Size
AT dropout (MSI sites tend to be GC-low)
Confirming Specificity
Below is a typical experimental design for assessing MSI specificity
Select a series of known MSS samples or tissue-matched non-tumor samples (e.g., not blood/buffy-coat samples if intended use is FFPE) spanning a typical range of sample qualities. Note that these samples should not be in your MSI panel-of-normals.
Run DRAGEN MSI on this sample set using your custom MSI sites file and using either the MSI reference files available here or constructing your own MSI reference files. For more information see DRAGEN MSI
Verify the range of Percentage Unstable Sites on these samples. It is expected that Percentage Unstable Sites should be low (ideally well below any established MSI threshold)
Confirming Sensitivity
Below is a typical experimental design for assessing MSI sensitivity
Select a series of known MSI-H samples spanning a range of typical sample qualities and expected Percentage Unstable Sites values, ideally with replicates. If needed, expected Percentage Unstable Sites values can be varied by diluting MSI-H samples into a matched normal sample, ideally matching sample types (e.g., FFPE normal not buffy coat normal if using FFPE tumor). If insufficient samples are available sample quality typically has less impact on sensitivity than on specificity so the use of cell lines may be an option.
Run DRAGEN MSI on this sample set using the same parameters as under Confirming Specificity.
Determine an MSI threshold value that reasonably separates MSI-H samples from the specificity cohort (in various Illumina solutions the value is typically between 10-20)
Characterize the precision of Percentage Unstable Sites values near this threshold (ideally using replicates) to establish approximate C5 and C95 values. C5 is the MSI score at which negative samples will be considered positive 5% of the time and C95 is the MSI score at which positive samples will be considered positive 95% of the time.
Last updated
Was this helpful?