Minimal Checklist
This section provides a succinct checklist of critical Quality Control (QC) metrics for evaluating DRAGEN run performance. These metrics are located in the standard CSV output files in the run directory.
1. Basic Run & Alignment QC
Source File: <output_prefix>.mapping_metrics.csv
Default Status: Enabled
Note that mapping metrics are computed on the raw input sample (prior to optional UMI collapsing).
Total input reads
Volume check: Confirm read count matches sequencer output and assay expectations. E.g. ~600–900M reads for 30X human WGS.
Q30 bases
Base quality: Generally > 85–90%. A global drop indicates chemistry or flow cell issues.
Mapped reads
Alignment success: Aim for > 95% (human WGS). Low mapping (< 90%) suggests wrong reference, contamination, or poor library quality.
Soft-clipped bases (R1/R2)
Adapter/quality trimming: Typical values < 2–3% for high-quality libraries. Values > 5% indicate adapter read-through, short inserts, or degraded DNA.
Estimated sample contamination
Purity check: Requires --qc-detect-contamination=true. For germline samples, > 1–2% can impact variant calling accuracy.
Duplicate reads
Library complexity: Germline WGS typically < 20%. WES and targeted assays often show 20–50%, which is expected given the reduced target space.
Properly paired reads
Fragment integrity: Should be > 90% for high-quality libraries. Low values indicate fragment issues or reference mismatch. RNA-seq expects lower rates due to splicing.
Reads with MAPQ [40:inf)
High-confidence alignments: For human WGS, expect > 80% with MAPQ ≥ 40. Lower values suggest alignment ambiguity or wrong reference.
2. FastQC (Sequence Composition)
Source File: <output_prefix>.fastqc_metrics.csv
Default Status: Enabled
Note that FastQC metrics are computed on the raw input sample (prior to optional UMI collapsing).
Positional base mean quality
Cycle decay: Identify quality drop-off at read ends or specific cycles indicating fluidics issues.
Read GC content
Contamination/bias: Deviation from expected distribution (e.g. human ~40–45% GC) suggests contamination or severe PCR bias.
Sequence positions (adapters)
Adapter content: High levels indicate untrimmed adapters or short inserts, leading to elevated soft-clipped bases in mapping metrics.
3. UMI QC (Applies Only to UMI Designs)
Source File: <output_prefix>.umi_metrics.csv
Default Status: Enabled as part of the UMI pipeline (requires --umi-enable=true)
Consensus reads
Conversion efficiency: Low ratio relative to input reads suggests low molecular complexity or insufficient sequencing depth.
Mean family size
Saturation: Family sizes near 1.0 indicate under-sequencing. Very high sizes may indicate excessive PCR amplification.
4. Coverage QC
Source File: <output_prefix>.*_coverage_metrics.csv
(e.g. wgs_coverage_metrics.csv or target_bed_coverage_metrics.csv)
Default Status: Enabled
Please note coverage metrics are computed post read deduplication or UMI collapsing. Reads with MAPQ=0 are ignored.
Average alignment coverage over target bed / genome
Depth check: Primary driver of sensitivity. Ensure it meets assay target (e.g. 30× germline, 100×+ somatic).
Uniformity of coverage (PCT > 0.2× mean)
Bias check: Low uniformity indicates coverage bias (e.g. GC bias), leading to variant calling blind spots. Germline WGS typically ≥ 80–90%.
PCT of genome with coverage [x: inf)
Callability: E.g. PCT ≥ 20×. Germline WGS typically requires > 95% at 20×.
Aligned bases in target bed / genome
Yield: Total usable data for normalizing performance across runs or flow cells.
5. Variant-Level Sanity Check (Optional)
Source File: <output_prefix>.vc_metrics.csv
Default Status: Conditional (variant caller enabled)
Ti/Tv Ratio
Biological plausibility: For human germline WGS, expect ~1.9–2.2. Values < 1.7 often indicate elevated false-positive rates.
Note: Ti/Tv is a biological sanity check, not a standalone QC pass/fail metric. Expected values depend on organism, assay type, and genomic region.
How to Use This Checklist
Treat this as a minimum QC review, not an exhaustive list.
Values outside these ranges warrant investigation but are not automatic failures.
Some metrics are assay-specific and may require optimization for particular use cases.
Last updated
Was this helpful?